Install high and low temperature tester should pay attention to those matters

Small-scale high-low temperature testing machine is a reliability testing equipment dedicated to R&D and production by Qinzhuo brand professional. It upholds professional technical services. The following notes on the installation of small-scale high-low temperature test boxes are explained as follows:


Small high and low temperature testing machine
1. When the small-sized high-low temperature testing machine is being transported, the inclination angle shall not be greater than 45°C to avoid damage to the refrigerator.
2, small high-low temperature testing machine to avoid violent shock and impact in order to avoid glass door fragmentation.
3, small high-low temperature testing machine should be far away from the source of electromagnetic interference, and should be grounded to the ground of the socket supplying the light box power supply.
4, small high-low temperature testing machine to be installed in places not exposed to direct sunlight, there must be adequate ventilation.
5, small high-low temperature testing machine to be installed on a solid and flat ground, and away from heat and the surrounding environment is not prone to high humidity.
6. Do not adjust the humidity setting too low to avoid constant dehumidification.
7. The humidifier tank must use distilled water or pure water. The amount of water must not be less than 1/3 of the volume. When water is added, care should be taken not to spill it. If it overflows or spills, wipe it with a rag to prevent water from flowing into the equipment and causing damage.
8. Axial-flow fan is located directly above the working chamber. Do not block the fan when it is placed to avoid uneven ventilation caused by poor ventilation.
9. Before the equipment is idle, remove the water from the humidifier tank and let the moisture inside the equipment completely dissipate. Before closing the door, confirm that the inside of the box is completely dry.
10. Before the equipment is moved, the water in the humidifying pan should be emptied. Spilled or spilled water can cause electrical leakage or electric shock.
11. Do not wash the device with solvents or other halide solutions as this may cause rust.
12. If no humidification is required, set the humidity to 0% RH and clean the water in the tank.
13, small high-low temperature tester If the ambient temperature is low and set the temperature is high, the measured temperature can not reach the desired set value for a long time, please turn the cooling capacity switch on the panel to the low position.